Fourier transform infrared (FTIR) spectroscopy can be used to determine the concentration and speciation of dissolved water in silicate glasses if the molar absorptivity coefficients (ε) are known. Samples that are thin and/or water-poor typically require the use of the mid-IR 3500 cm–1 total water (H2Ot) and 1630 cm–1 molecular water (H2Om) absorbance bands, from which hydroxyl water (OH) must be determined by difference; however, accurate determination of H2Ot and OH is complicated because ε3500 varies with water speciation, which is not usually known a priori. We derive an equation that uses end-member ε3500 values to find accurate H2Ot and OH concentrations from the 3500 cm–1 absorbance for samples where only the H2Om concentration need be known (e.g., from the 1630 cm–1 band). We validate this new species-dependent ε3500 method against published data sets and new analyses of glass standards. We use published data to calculate new end-member ε3500 values of ε3500OH = 79 ± 11 and ε3500H2Om = 49 ± 6 L/mol·cm for Fe-bearing andesite and ε3500OH = 76 ± 12 and ε3500H2Om = 62 ± 7 L/mol·cm for Fe-free andesite. These supplement existing end-member values for rhyolite and albite compositions. We demonstrate that accounting for the species-dependence of ε3500 is especially important for hydrated samples, which contain excess H2Om, and that accurate measurement of OH concentration, in conjunction with published speciation models, enables reconstruction of original pre-hydration water contents. Although previous studies of hydrous silicate glasses have suggested that values of ε decrease with decreasing tetrahedral cation fraction of the glass, this trend is not seen in the four sets of end-member ε3500 values presented here. We expect that future FTIR studies that derive end-member ε3500 values for additional compositions will therefore not only enable the species-dependent ε3500 method to be applied more widely, but will also offer additional insights into the relationship between values of ε and glass composition.

The dissolved water content of a silicate melt affects a range of magmatic processes, since it exerts a strong influence on melt viscosity (e.g., Hess and Dingwell 1996; Giordano et al. 2008), volatile diffusivities (e.g., Baker et al. 2005; Baker and Alletti 2012), and crystallization (e.g., Hammer 2004; Gualda et al. 2012). Accurate measurement of the water content of silicate glasses is therefore crucial to a wide range of volcanological and petro-logical studies. Additionally, measuring water speciation, i.e., the amount of water present as molecular water (H2Om) vs. hydroxyl groups (OH), can provide information about the cooling rate and glass transition temperature (Tg) of a sample (e.g., Stolper 1989; Dingwell and Webb 1990; Xu and Zhang 2002), and whether a sample has been affected by hydration during bubble resorption, or crystallization during cooling (e.g., McIntosh et al. 2014; Nichols et al. 2014), or by secondary hydration at low temperature in the time following deposition (e.g., Dixon et al. 1995; Anovitz et al. 2008; Denton et al. 2009; Tuffen et al. 2010). Whereas most commonly used techniques [e.g., hydrogen manometry, thermogravimetric analysis (TGA), secondary ion mass spectrometry (SIMS)] can measure only total water (H2Ot) contents, Fourier transform infrared spectroscopy (FTIR) is capable of measuring both total water and water species concentrations.

During transmission FTIR analysis, some infrared light passing through the sample is absorbed by molecules within the sample, with different species absorbing light at different, characteristic frequencies. The resulting FTIR spectrum thus exhibits absorbance bands at wavelengths (or wavenumbers, cm–1) corresponding to these different absorbing species. There are four main absorbance bands associated with water dissolved in silicate glasses that can be seen in the near- and mid-IR range (Fig. 1), and their assignments are given in Table 1. The H2Om and OH bands at 5200 and 4500 cm–1 have low absorption intensities and are often not detected in glasses that have low water contents or that require thin samples for analysis (e.g., to avoid bubbles or crystals in the beampath). Additionally, these bands can be difficult to measure accurately in intermediate glass compositions where absorptions by Fe cause curvature of the baseline at high wavenumbers (e.g., Stolper 1982b; Ohlhorst et al. 2001; Mandeville et al. 2002), or for samples where the absorptions creating the 4500 cm–1 band may be complex (Malfait 2009; Le Losq et al. 2015). In many situations therefore the 3500 cm–1 band is used to measure the total water (H2Ot) concentration, and water speciation is obtained by measuring the 1630 cm–1 H2Om band and then finding OH indirectly by difference (e.g., Wysoczanski and Tani 2006; Yokoyama et al. 2008; Nichols et al. 2009; McIntosh et al. 2014; Giachetti et al. 2015).

The intensity of the absorbance bands is typically measured as the height of the absorbance peak above the baseline and is termed the absorbance. These absorbance values are then converted into concentrations via the Beer-Lambert law, which relates the attenuation (absorbance) of light passing through a material to the thickness of the material and the concentration of the absorbing species within it, according to:
(1)
where C is the concentration of the species of interest (CH2Ot, CH2Om, or COH in wt%), M is the molecular weight (in g/mol; 18.02 for water), A is the absorbance (no units), ρ is the density (in g/L), l is the thickness of the analyzed area (in cm), and ε is the molar absorptivity coefficient for the absorbance band of interest (in L/mol·cm). If the intensity of the absorbance band is measured using the area enclosed by the peak and the baseline rather than the peak height, then this is termed the integrated absorbance (in cm–1) and ε in Equation 1 is replaced with ε*, the integrated molar absorptivity (in L/mol·cm2). Since the value of the peak area is more sensitive to errors in fitting the baseline than the peak height, it is more common to use the absorbance rather than the integrated absorbance. Values of ε (and ε*) are experimentally determined, and, for a given water absorbance band will vary with the anhydrous glass composition, such as from basalt to rhyolite (Silver et al. 1990; Dixon et al. 1995; Ohlhorst et al. 2001; Mandeville et al. 2002; Seaman et al. 2009; Mercier et al. 2010). Establishing the correct value of ε to use for a given species in glass of a given composition is therefore fundamental to the quantitative use of FTIR spectroscopy.

It has been previously noted that, because the 3500 cm–1 absorbance band is the net result of absorptions by both H2Om and OH species (Table 1), the correct molar absorptivity coefficient for that band—termed ε3500—will vary with the ratio of the two species (Newman et al. 1986; Okumura et al. 2003). Since the equilibrium proportions of water species vary with temperature and total water concentration (e.g., Stolper 1982a, 1989), and can be altered by disequilibrium processes such as quench resorption (McIntosh et al. 2014), crystallization (e.g., Nichols et al. 2014), and secondary hydration (e.g., Dixon et al. 1995; Anovitz et al. 2008; Tuffen et al. 2010), the “true” ε3500 value will also vary accordingly, potentially even across the same sample. This accounts for the wide range in ε3500 values reported in the literature, even among glasses of the same composition. For example, rhyolites, which are the most studied composition to date, have reported ε3500 values ranging from 75 to 95 L/mol·cm (Table 2).

Newman et al. (1986) show that end-member molar absorptivity coefficients can be calculated for the 3500 cm–1 band; these give the theoretical ε3500 value if all water within the glass were present exclusively as H2Om (ε3500H2Om), or exclusively as OH (ε3500OH). They demonstrate that these coefficients can be used to determine an accurate species-dependent ε3500 value, according to:
(2)
where XOH and XH2Om are the simple mass fractions of water dissolved as OH or H2Om, respectively. For their data set of hydrous rhyolite compositions, Newman et al. (1986) calculate that ε3500OH = 100 ± 2 and ε3500H2Om = 56 ± 4 L/mol·cm. However, as the authors point out, this simple relationship for calculating a species-dependent ε3500 value is of limited analytical use since it requires that the speciation of the sample is already known. This requirement has to date limited the practical use of a species-dependent ε3500 value and many researchers have instead necessarily, albeit somewhat arbitrarily, selected a constant value from the literature. Here we build on the work of Newman et al. (1986) and present a new methodology for accounting for the species-dependence of ε3500, without requiring a priori knowledge of the species proportions. This method enables accurate water species concentrations to be determined from the 3500 and 1630 cm–1 absorbance bands. We demonstrate the effectiveness of this technique by applying it to published data sets and to the analyses of glass standards with known water contents, and then discuss its implications for the analysis of hydrated glasses.

Calculation of OH concentration using end-member ε3500 values

Equation 2 can be rewritten in terms of species concentrations, using the definitions XOH = COH/CH2Ot and XH2Om = CH2Om/CH2Ot, giving
(3)
Rearranging for CH2Ot, and substituting into the Beer-Lambert law (Eq. 1), we obtain an expression for the concentration of OH
(4)
where A3500 is the measured 3500 cm–1 absorbance normalized for sample thickness (i.e., A = A/l) in units of 1/cm. With this equation it is now possible to calculate directly the OH concentration of a sample if the glass thickness, 3500 cm–1 absorbance, H2Om concentration (which can be found from either the 5200 or 1630 cm–1 absorbance band in the conventional way via Eq. 1), and the end-member ε3500 values for the glass composition of interest are known. H2Ot concentration is then simply CH2Ot = COH + CH2Om.

To test the accuracy of this new species-dependent ε3500 method we apply it to the published data set of Newman et al. (1986) from which they calculated their end-member ε3500 values for rhyolite. Their data set consists of samples of natural rhyolitic obsidian from tephra deposits, domes, and flows taken from a range of locations in the U.S.A. Thickness data and 5200, 4500, and 3500 cm–1 absorbances are reported for 28 analyses of 24 different samples, containing 0.27–2.64 wt% H2Ot as measured by manometry. For these samples, we calculate their H2Ot and OH concentrations in the conventional way (Eq. 1) using the published 5200 cm–1 (H2Om) and 4500 cm–1 (OH) absorbance data together with the ε5200 and ε4500 values for rhyolite derived by Newman et al. in the same study (Table 2). These are then plotted against the H2Ot and OH concentrations of the same samples calculated from the published 3500 cm–1 (H2Ot) and 1630 cm–1 (H2Om) absorbances using our new species-dependent method with the values of ε3500OH and ε3500H2Om for rhyolite derived by Newman et al. (Fig. 2). We find that there is excellent agreement between H2Ot and OH concentrations calculated by the two methods, as demonstrated by the excellent fit to the 1:1 line (R2 = 0.993 for H2Ot, and R2 = 0.971 for OH). The “true” (species-dependent) ε3500 values for individual samples then calculated via Equation 2 range from 73.6 to 97.2 L/mol·cm, compared to the end-member values of ε3500OH = 100 ± 2 and ε3500H2Om = 56 ± 4 L/mol·cm.

Calculating end-member ε3500 values for different compositions

New end-member ε3500 values can be calculated for different glass compositions following the procedure of Newman et al. (1986), who showed that the absorbances of the 5200, 4500, and 3500 cm–1 bands and their molar absorptivity coefficients can be related as:
(5)
This has the form y = m1x1 + m2x2, thereby enabling the use of multiple linear regression (e.g., using the Linest function in Microsoft Excel) to find the values of ε3500H2Om/ε5200 and ε3500OH/ε4500 from the measured 5200 and 4500 cm–1 absorbances. Values of ε5200 and ε4500 reported in the literature can then be used to find the two unknowns, i.e., the end-member coefficients ε3500H2Om and ε3500OH.

Thus to calculate end-member ε3500 values for a given glass composition, it is necessary to have a data set containing 5200, 4500, and 3500 cm–1 absorbance and thickness data for the same piece(s) of glass, for which the ε5200 and ε4500 values are also known. Since the absorbances are normalized to thickness it is possible to measure the weaker 5200 and 4500 cm–1 absorbances on a thicker piece of glass and then subsequently thin it to measure the 3500 cm–1 absorbance. Additionally, it would also be possible to use the 1630 cm–1 H2Om absorbance and ε1630 value in place of the 5200 cm–1 H2Om absorbance and ε5200 value, if necessary.

We demonstrate this procedure by using the published data set of Mandeville et al. (2002) to calculate new end-member ε3500 values for both their Fe-bearing and Fe-free andesite compositions. Their samples are hydrous glasses that were synthesized at high pressure and temperature from either a mixture of natural basaltic andesite and evolved andesite rock powders from Krakatau, or a mixture of pure oxide powders and carbonates; H2Ot contents were measured by hydrogen manometry (see original study for further details). In their data set, glasses with data for all three bands (5200, 4500, and 3500 cm–1) include two Fe-bearing andesite glasses, Run 9 (1030 °C, 200 MPa, 4.32 wt% H2Ot, 28 analyses) and Run 101 (1100 °C, 50 MPa, 1.67 wt% H2Ot, 13 analyses), and three Fe-free andesite glasses, Run 58 (1100 °C, 200 MPa, 5.68 wt% H2Ot, 56 analyses), Run 68 (1100 °C, 150 MPa, 2.15 wt% H2Ot, 11 analyses), and Run 106 (1050 °C, 65 MPa, 1.31 wt% H2Ot, 9 analyses). From the normalized absorbances of these glasses we used the Linest function in Microsoft Excel to find values of ε3500H2Om/ε5200 and ε3500OH/ε4500, as per Equation 5, for each composition, from which the end-member ε3500OH and ε3500H2Om coefficients were calculated using the values of ε5200 and ε4500 previously derived and published in the original study (Table 2).

For the Fe-bearing andesite glass compositions reported in Mandeville et al. we thus calculate end-member ε3500 values of ε3500OH = 79 ± 11 and ε3500H2Om = 49 ± 6 L/mol·cm, while for their Fe-free andesite composition we calculate values of ε3500OH = 76 ± 12 and ε3500H2Om = 62 ± 7 L/mol·cm. Errors were calculated by propagating the standard error on the regression coefficients and the reported error on the ε5200 and ε4500 values used to derive them via Equation 5. Figure 3 compares the H2Ot and OH concentrations calculated from the 5200 and 4500 cm–1 bands with those calculated from the 3500 and 1630 cm–1 bands using both the species-dependent ε3500 method (Eq. 4) with the new end-member ε3500 values (filled symbols) and the conventional method (Eq. 1) with the fixed ε3500 value derived by Mandeville et al. (open symbols). For both compositions, the H2Ot concentrations calculated using the species-dependent ε3500 method are in slightly better agreement with the H2Ot concentrations calculated from the 5200 and 4500 cm–1 bands than the data calculated using the fixed ε3500 value (Figs. 3a and 3c). Meanwhile, the OH concentrations calculated using the species-dependent ε3500 method are in markedly better agreement with the OH concentrations calculated from the 4500 cm–1 band than those calculated from the 3500 cm–1 band using the conventional OH-by-difference method. For Fe-bearing andesite the OH-by-difference method gives no correlation with OH concentrations derived directly from the 4500 cm–1 band (R2 = 0.10), whereas the new method gives a much better fit (R2 = 0.80; Fig. 3b). The OH data for Fe-free andesite are somewhat more scattered with regards to the 1:1 line, with most of the scatter corresponding to one glass (Run 58a), but the new method again is closer to the gradient of the 1:1 line and provides a better correlation than the OH-by-difference method, with R2 values of 0.66 and 0.55, respectively (Fig. 3d).

A previous study by Silver and Stolper (1989) of H2O in albitic glasses also followed the method of Newman et al. (1986) in determining end-member ε3500 values for the 3500 cm–1 band, but since they were similar they elected to publish a single fixed ε3500 value instead. For completeness, we use the published Silver and Stolper data set (absorbance data for 5 glasses, synthesized at 1500–2000 MPa and 1400–1600 °C with 1.02–5.12 wt% H2Ot) to calculate that the end-member ε3500 values would be ε3500OH = 69 ± 17 and ε3500H2Om = 71 ± 17 L/mol·cm, compared to the published fixed ε3500 value of 70 ± 2 L/mol·cm (Table 2).

Application to hydrous glass standards

To further test the accuracy of Equation 4 we performed transmission FTIR analyses of three glass standards with independently constrained water contents (NWC, KRA-045-2, Run 10). Chemical compositions and densities of these glasses are reported in Table 3. NWC is a rhyolitic obsidian from the Upper Dome of NW Coulee at Mono Craters, California. A piece of this glass was included in the data set of Hauri et al. (2002) (“NW Coulee”), whose manometry analysis found it to contain 0.297 wt% H2Ot. KRA-045-2 is a rhyodacitic obsidian matrix glass from the pyroclastic flow deposit of the 1883 Krakatau eruption (Mandeville et al. 1998), which was analyzed by manometry and found to contain 0.48 wt% H2Ot [C. Mandeville, published in Maria and Luhr (2008)]. Finally, Run 10 is an experimentally synthesized Fe-bearing andesite glass that was part of the data set of Mandeville et al. (2002), whose manometry analysis gives its total water content as 5.78 wt% H2Ot.

All three glasses were prepared for transmission FTIR analysis as double-polished free-standing wafers. The glass thicknesses of the rhyolite standards (NWC and KRA-045-2) were measured using a digital micrometer with a precision of ±1 μm, with measurements taken at the location of each analysis. Run 10 was too thin, hence fragile, to measure with the micrometer and so thickness was instead calculated from fringes in reflectance spectra following the method of Wysoczanski and Tani (2006) and utilizing a refractive index of 1.55 for andesite (Wohletz and Heiken 1992), with an estimated accuracy of ±3 μm. FTIR spot analyses were performed at the Japan Agency for Marine-Earth Science and Technology, using a Varian FTS 7000 spectrometer and an attached UMA600 microscope. Mid-IR (6000–700 cm–1) transmittance spectra were collected over 512 scans at a resolution of 8 cm–1 using a heated ceramic IR source, a KBr beamsplitter and a liquid nitrogen-cooled HgCdTe2 (MCT) detector. Apertures were used to reduce the beam spot size to 20 × 20 μm. Both the spectrometer bench and microscope were continuously purged with N2 gas to minimize any interference from atmospheric H2O. Obtained spectra were then processed using Win-IR Pro software. Band absorbances were determined by measuring the height of the peak above the baseline. For the 3500 cm–1 band a linear baseline was used, while for the 4500 and 1630 cm–1 bands baselines were drawn by hand using a flexicurve. H2Om concentration was calculated from the 1630 cm–1 absorbance via the Beer-Lambert law (Eq. 1) using ε1630 values of 55 ± 2 L/mol·cm for rhyolite (Newman et al. 1986) and 42.34 ± 2.77 L/mol·cm for Fe-bearing andesite (Mandeville et al. 2002). OH concentration was then calculated according to Equation 4, utilizing the end-member ε3500 values of Newman et al. (1986) for rhyolite and the newly derived end-member ε3500 values for Fe-bearing andesite. H2Ot was then calculated as CH2Ot = COH + CH2Om. For comparison, CH2Ot and COH (using OH-by-difference) were also calculated from the 3500 cm–1 band via Equation 1 using the range of published fixed ε3500 values (Table 2). For NWC and KRA-045-2 OH concentration was also calculated via Equation 1 from the 4500 cm–1 band using appropriate values of ε4500 (Table 2).

Table 4 and Figures 4 and 5 show the results of the FTIR spot analyses of the two rhyolite standards, NWC and KRA-045-2, and the Fe-bearing andesite Run 10. Reported values for each sample are the mean value of 12 measurements made on the same wafer. Errors for the rhyolite standards are given as one standard deviation and for the Fe-bearing andesite are derived from the ±3 μm error on the thickness measurement, which becomes the dominant source of error for thin samples. For each glass, the H2Ot concentration measured by FTIR using both the species-dependent ε3500 method and the conventional fixed ε3500 method is plotted against their known H2Ot content previously measured by manometry (Figs. 4a and 5a). For all compositions, the species-dependent ε3500 method (filled squares) gives H2Ot concentrations that are within 5% of the known value, whereas those calculated using fixed ε3500 values from the literature (open squares) span a wide range and differ from the known H2Ot concentration by as much as 33%.

OH concentrations calculated using the species-dependent ε3500 method and the conventional OH-by-difference method are compared to the OH concentration obtained from the 4500 cm–1 band (Figs. 4b and 5b). For each glass, H2Ot concentrations determined by summing the 4500 cm–1 OH and 1630 cm–1 H2Om concentrations were compared to their manometry H2Ot concentration (Fig. 4c), with the fit to the 1:1 line used to select the most appropriate ε4500 value from the literature (Table 2). Since the weak 4500 cm–1 band could not be measured in our thin Run 10 standard we use instead the 4500 cm–1 absorbance data reported by Mandeville et al. (2002) for the same glass, and the ε4500 value derived in the same study (Table 2). For the rhyolite standards (Fig. 4b) the OH concentrations calculated by the species-dependent ε3500 method (filled circles) are within 12% of the 4500 cm–1 OH concentration for NWC and within 2% for KRA-045-2, whereas the OH-by-difference concentrations using fixed ε3500 values (open circles) differ by up to 45% and 27%, respectively. For the Fe-bearing andesite Run 10 (Fig. 5b), the species-dependent ε3500 method gives an OH concentration within 9% of that reported by Mandeville et al. (2002). Finding OH-by-difference using the fixed ε3500 value derived by Mandeville et al. (2002) (for the data set that contained Run 10) matches the 4500 cm–1 value exactly, whereas the OH-by-difference derived using the King et al. (2002) fixed ε3500 coefficient is 39% lower than the 4500 cm–1 value.

Accuracy of H2Ot and OH concentrations from species-dependent ε3500 method

The new species-dependent ε3500 method is able to reproduce accurately the H2Ot and OH concentrations of both the published rhyolite and andesite data sets (Figs. 2 and 3) and the hydrous glass standards (Figs. 4 and 5; Table 4). For the data sets, the H2Ot and OH concentrations derived from the 5200 cm–1 H2Om and 4500 cm–1 OH absorbance bands are not strictly independent measurements since they need ε5200 and ε4500 values to find concentration via the Beer-Lambert law. However, by using the ε5200 and ε4500 values derived from these same data sets in the original studies of Newman et al. (1986) and Mandeville et al. (2002) the resulting concentrations are particularly reliable. The H2Ot and OH concentrations from the new species-dependent ε3500 method have an excellent correlation with the values derived from the 5200 and 4500 cm–1 bands, demonstrating the accuracy of this technique (Figs. 2 and 3). For the andesite data sets, the species-dependent ε3500 method gives a better fit than the conventional method using fixed ε3500 values. Although the improvement for H2Ot is only slight, it is particularly marked for OH, where the species-dependent ε3500 method gives much better agreement with the 4500 cm–1 OH concentrations than the conventional OH-by-difference method. In particular, the OH-by-difference values for Fe-bearing andesite (Fig. 3b) give no meaningful correlation to the 4500 cm–1 OH concentrations (R2 = 0.10) while the species-dependent method gives a much stronger correlation (R2 = 0.79). We therefore strongly recommend the use of the species-dependent ε3500 method for finding the OH concentration when it is necessary to use the 3500 absorbance band to do so.

Similarly, FTIR analysis of the three glass standards using the species-dependent ε3500 method produces H2Ot and OH concentrations that are within 11% of their independently determined concentrations. The absolute difference is greatest for the Fe-bearing andesite Run 10, where the species-dependent ε3500 method overestimates H2Ot by 0.11 wt% (Fig. 5a). This discrepancy is most likely due to the need to analyze a thin glass wafer (to avoid saturation of the 3500 cm–1 band in this water-rich sample), which increases the relative error owing to the thickness measurement and also creates interference fringes, which introduce uncertainty when picking the baseline. Since Run 10 is water-rich however, this error is <2% of the total H2Ot concentration measured by manometry. By contrast, the H2Ot concentration calculated for the rhyolite NWC (Fig. 4a) is only 0.015 wt% higher than the manometry value, but due to the low H2Ot content of this sample the relative difference is higher (5%). Possible sources of error include slight heterogeneity in water content between the chip that was analyzed with manometry and the chip that was analyzed with FTIR, and differences in how the baseline was picked for the 1630 cm–1 H2Om peak (see below). Despite these minor discrepancies, the H2Ot and OH concentrations calculated using the species-dependent ε3500 method are markedly closer to the independently determined concentrations than some of the concentrations calculated using fixed ε3500 values from the literature, which may differ by up to 45% of the known values. Since using a fixed ε3500 value necessarily implies that OH concentration must be found as OH-by-difference, the resulting errors in H2Ot concentration are also propagated through as errors in OH concentration.

Accuracy of H2Om measurement

Since the H2Om concentration is needed to calculate the species-dependent ε3500 value, any error in the H2Om concentration affects the calculated OH and H2Ot concentrations. When using the 3500 cm–1 band, the H2Om concentration will generally be determined from the 1630 cm–1 band. The ε1630 value for rhyolite comes from Newman et al. (1986) (Table 2), who used the spectrum of an anhydrous equivalent of the glass to define the baseline when measuring 1630 cm–1 absorbances, in recognition of the silicate peak at ~1600 cm–1 that can interfere with the 1630 cm–1 H2Om peak and may cause the 1630 cm–1 absorbance to be overestimated if using a linear baseline. In lieu of an anhydrous piece of the same glass, we used a hand-drawn flexicurve baseline pinned to the baseline immediately adjacent to the 1630 cm–1 peak to account for non-linearity of the baseline. This flexicurve baseline gives an H2Om concentration of 0.04 wt% (and subsequent H2Ot concentration of 0.31 wt%), compared to 0.06 wt% H2Om (and 0.32 wt% H2Ot) if a linear baseline is used. The good agreement of our flexicurve H2Om concentration measurement with the H2Om concentrations measured by Newman et al. (1986) for NWC (0.05 wt% using 1630 cm–1 band and 0.03 wt% using 5200 cm–1 band) validate our method of fitting the 1630 cm–1 baseline, and the small difference between our FTIR H2Ot concentration value and the manometry value (0.015 wt%) indicates that its impact is limited. The weak intensity of the ~1600 cm–1 silicate peak means that significant absorbance is restricted to thick sample wafers, while the relative importance of its interference with the 1630 peak is greatest for samples with low H2Om concentrations (McIntosh et al. 2015). The error introduced by the 1630 cm–1 baseline will therefore be greatest for water-poor samples, but since these samples should contain predominantly OH the ultimate error on the calculated OH and H2Ot concentrations using the species-dependent ε3500 method will be limited.

The accuracy of the H2Om measurement will also necessarily rely on the accuracy of the ε1630 value used to obtain it. When using the 3500 and 1630 cm–1 bands along with a fixed ε3500 value to find OH concentration as OH = H2Ot – H2Om, the resulting error for the OH concentration may previously have been attributed to an inaccurate value of ε1630, rather than ε3500. We expect that the use of the species-dependent ε3500 method will reduce or remove such apparent discrepancies in H2O species concentrations, and that the existing literature values for ε1630 in the compositions discussed in this paper will prove to be robust. It is worth noting that there is evidence that the ε values of the 5200 and 4500 cm–1 bands can vary with H2Ot concentration (Zhang et al. 1997b; Yamashita et al. 2008). The excellent fit of the H2Ot and OH concentrations calculated using the species-dependent ε3500 method to the published Newman rhyolite and Mandeville andesite data sets, which span a wide range of H2Ot concentrations, supports our assumption that the ε3500H2Om and ε3500OH values are true end-member values and do not themselves vary with H2Ot concentration. It is possible however that in the future a similar H2Ot-dependence will be found for ε1630 in some compositions, in which case we expect that our species-dependent ε3500 method can be adapted to account for it. As an example, we show that the H2Ot-dependent relationship derived for the 5200 cm–1 band by Zhang et al. (1997) can be successfully substituted into our method in place of the 1630 cm–1 band (see Supplementary1 spreadsheet). In cases where both the 5200 and 1630 cm–1 bands can be measured for the same glass, we recommend using H2Om concentrations calculated from both as input into Equation 4 to check and quantify possible error stemming from the H2Om measurement.

Comparison of species-dependent ε3500 and fixed ε3500 values in literature

Since the true ε3500 value is species-dependent, the range in ε3500 values published in the literature results from the differences in the overall speciation (i.e., the ratio of OH to H2Om) of the samples used in the different calibration studies. In silicate melts and glasses, the two species interconvert according to the equilibrium reaction:
(6)
in which H2Om reacts with bridging O atoms (O°) to produce OH groups that are bound to the silicate framework (Stolper 1982a). For a given melt composition, the position of this equilibrium reaction (i.e., equilibrium speciation) is controlled by the temperature and H2Ot content (e.g., Stolper 1982a, 1989; Nowak and Behrens 1995; Behrens and Nowak 2003; Behrens and Yamashita 2008), with the equilibrium shifted toward OH at high temperature and low H2Ot, and toward H2Om at low temperature and high H2Ot (Fig. 6). Thus the true value of ε3500 will shift toward the higher ε3500OH end-member value for high temperatures and/or low H2Ot concentrations, and toward the lower ε3500 H2Om end-member value for lower temperatures and/or high H2Ot concentrations. A final consideration is that the species interconversion reaction rate is strongly temperature dependent and slows dramatically during cooling (Zhang et al. 1991, 1995), until interconversion becomes negligible and speciation becomes “frozen in.” The temperature at which this occurs is termed the temperature of apparent equilibrium, Tae, and for a given H2Ot content will be higher for a fast quench rate (since less time is available at every temperature interval for the reaction to approach equilibrium), and lower for a slow quench rate (e.g., Zhang et al. 1995; Xu and Zhang 2002). Tae has been shown to be equivalent to the temperature of the glass transition, Tg (Dingwell and Webb 1990; Zhang et al. 1997a). However, since Tg is most strictly defined as the glass transition temperature for a melt under a particular set of experimental conditions (e.g., a cooling rate of ~10 K/min), for clarity we use Tae in the following discussion of samples with varied cooling histories. Thus the speciation of rapidly quenched samples will reflect their high Tae by having higher COH:CH2Om ratios, hence higher ε3500 values, than equivalent samples with slower quench and lower Tae.

The variation in published ε3500 values for a given glass composition therefore results from differences in both the water contents and temperature histories of the samples used in each calibration study. For Fe-bearing andesite, Mandeville et al. (2002) derived their fixed ε3500 value of 62.32 ± 0.42 from four glasses synthesized at 1030–1130 °C containing 0.21–4.32 wt% H2Ot, with 39 of their 63 analyses stemming from their Run 9 glass (1030 °C, 4.32 wt% H2Ot), while King et al. (2002) derived a fixed ε3500 value of 70.3 ± 6.86 using glasses synthesized at 1300 °C and containing 0–6.09 wt% H2Ot, of which only three out of 43 analyses were made on glasses with >3.39 wt% H2Ot. The higher ε3500 value of King et al. (2002) therefore likely reflects the lower average H2Ot content of their samples and their use of a higher experimental temperature and rapid quench rate (with samples quenched to below the glass transition in ~4 s), all of which favor higher COH:CH2Om. Applying the fixed ε3500 value of King et al. (2002) to our measurements of the Fe-bearing andesite standard Run 10 underestimates the true H2Ot content of this glass by 0.72 wt% (Fig. 5a), highlighting the potential for large errors when applying a fixed ε3500 value to samples created under different conditions.

Our analyses of the rhyolite standards NWC and KRA-045-2 produced “true” ε3500 values of 94.7 and 93.1, respectively. These high ε3500 values reflect the dominance of OH at low H2Ot concentrations (Fig. 6), and are close to the highest fixed ε3500 values reported in the literature: 90 (Hauri et al. 2002) and 95 ± 8 (Aubaud et al. 2009). The calibration of Hauri et al. (2002) was based on a wide range of samples, the majority of which had <0.8 wt% H2Ot, while that of Aubaud et al. (2009) was based on two samples with <0.16 wt% H2Ot. At such low H2Ot contents only negligible H2Om would be expected (Fig. 6) and it is therefore reasonable that these fixed ε3500 values should approach the end-member ε3500OH value of 100 ± 2 for rhyolite. Although Leschik et al. (2004) found that for glasses with >2 wt% H2Ot the ε3500 value decreases with increasing H2Ot concentration (as would be expected due to the increasing proportion of H2Om and corresponding shift toward the lower ε3500H2Om end-member), it is unclear why their fixed ε3500 value of 80 ± 4.9 for water-poor, OH-dominated samples (like our standards) is so much lower than both the Newman et al. ε3500OH end-member and other fixed ε3500 values.

By contrast the lowest published value for a fixed ε3500 value for rhyolite (75 ± 4) is that of Okumura et al. (2003), and is based on obsidian samples with 0.24–1.25 wt% H2Ot that were either unheated or heated to 500–700 °C. The authors noted that ε3500 values derived for the same obsidian source and H2Ot content (OBSW, 0.74 wt% H2Ot) increased with the experimental temperature, being 73 for 500 °C, 76 for 600 °C, and 80 for 700 °C. As discussed by the authors, this variation can be explained by the temperature dependence of H2O speciation that favors higher COH:CH2Om, hence higher ε3500, at higher temperatures. Okumura et al. (2003) also derived an ε3500 value for the original unheated obsidian of 77 L/mol·cm. Based on their observed temperature dependence of ε3500, we suggest that this value implies that the original Tae of this obsidian was ~620 °C. While this indicates that finding the true ε3500 of a sample could be a useful method of finding its Tae (≈ Tg), we stress that it is only valid if the temperature dependence of ε3500 is known for samples with exactly the same H2Ot content. Any derivation of a fixed value of ε3500 is effectively unique to samples with the exact same H2Ot content and temperature history, hence water speciation, as the samples that were used in its original calibration. Provided that end-member ε3500 values exist for the glass composition of interest, the advantage of our species-dependent ε3500 method is that it is possible to obtain accurate H2Ot and OH concentrations regardless of a sample's H2Ot content or temperature history, and it can also account for changes in H2Ot concentration across an individual sample.

Application to hydrated samples

Volcanic glasses are susceptible to secondary hydration, i.e., the addition of water at low temperature in the time following eruption (Friedman and Smith 1958). Hydration of obsidian in particular has a long history of study, not only by the geological, but also the archaeological community, since diffusion modeling of hydration profiles at glass margins could offer a way to date obsidian flows or tools (e.g., Friedman and Long 1976, 1984; Anovitz et al. 1999; Riciputi et al. 2002). Recent studies have also demonstrated that secondary hydration is widespread and has altered the glass water contents of many erupted pyroclasts, with the effect most pronounced for samples with greater surface area exposed to outside water, such as vesicular glasses (e.g., Giachetti and Gonnermann 2013; Dingwell et al. 2015). Determining the original eruptive H2Ot content of hydrated glasses is therefore critical to volatile studies of erupted pyroclasts. Here we use an example of obsidian hydration from the literature to discuss how the species-dependent ε3500 method can be used to measure accurately the water species concentrations of hydrated glasses, with the potential to thereby reconstruct the original H2Ot contents of hydrated glasses.

Yokoyama et al. (2008) used transmission FTIR to analyze hydration profiles at the margins of two obsidian flows from Kozushima, Japan, dated to 26 000 and 52 000 yr before present. Using the 3500 and 1630 cm–1 bands (and thus finding OH-by-difference), their analyses showed that both H2Om and H2Ot concentrations increase toward the hydrated boundary. The trend in OH concentration was less simple, however, with OH either increasing or decreasing toward the boundary depending on the choice of the fixed ε3500 value. To prevent negative OH concentrations, they had to use a fixed ε3500 value of 60; much lower than any of the published values of ε3500 for rhyolite (Table 2). We extracted the concentration data from their published profiles and used their stated values of thickness and density to back-calculate their original 3500 and 1630 cm–1 absorbances using the Beer-Lambert law. We then applied our new species-dependent ε3500 method to their absorbances to recalculate the H2Ot and OH concentrations of their hydration profiles, and to calculate the “true” ε3500 value for every point along the profile (Fig. 7). Doing so, we find that for both profiles the ε3500 value decreases toward the hydrated boundary. Accordingly, the OH concentration profiles no longer exhibit a fall to negative values and remain within 0.1 wt% of their non-hydrated values, while the increase in H2Ot concentrations toward the boundary becomes more pronounced than in the original profiles.

Since the species interconversion reaction effectively stops at the glass transition temperature (e.g., Dingwell and Webb 1990; Nowak and Behrens 1995; Zhang 1999; Behrens and Nowak 2003; Behrens and Yamashita 2008), water added at ambient temperature during secondary hydration is added as H2Om and not interconverted to OH; this creates “disequilibrium” speciation similar to that which develops during quench resorption (McIntosh et al. 2014). Glass in the hydrated margin becomes enriched in H2Om, and the correct ε3500 value to use will shift toward the ε3500H2Om end-member value (56 ± 2 for rhyolites). This explains why Yokoyama et al. (2008) had to use an ε3500 value of 60 to prevent negative OH concentrations when finding OH-by-difference. This value is much lower than the fixed ε3500 values in the literature, since the literature values were derived from samples with equilibrium rather than H2Om-rich disequilibrium speciation. When measuring H2Ot and OH concentrations of hydrated samples using the 3500 cm–1 band, it is therefore imperative that the species-dependence of the ε3500 coefficient is accounted for.

The study of Yokoyama et al. (2008) was limited by the spatial resolution of their FTIR apparatus (they used 15 × 50 μm spots at overlapping 5 μm steps) and we therefore truncate our recalculated profiles at the distance from the edge at which the authors calculated there would be no interference from the adjacent resin. FTIR apparatus with higher spatial resolution, such as those using a synchrotron source (e.g., von Aulock et al. 2014), may be able to investigate the concentration variations in hydrated margins in more detail. Nevertheless, it is clear from the data of Yokoyama et al. (2008) that OH concentration does not increase sharply in the hydration rim, whereas the H2Om concentration does, supporting their conclusion that the dominant species diffusing into the glass is H2Om. A recent study (Bindeman and Lowenstern 2016) on hydration of rhyolite during perlite formation at Yellowstone concluded that hydration occurred at temperatures <200 °C but greater than ambient temperature, over an expected cooling timescale of weeks to years. The authors observed that, although dominated by addition of H2Om, this hydration also added minor amounts of OH (~0.2 wt%) to hydrated rims. Our recalculated hydration profile of Yokoyama et al.'s Ohsawa lava (Fig. 7a) reveals a slight increase in OH concentration from ~0.1 to ~0.2 wt% toward the margin, raising the possibility that its hydration may have occurred under similar conditions to that of the Yellowstone perlite. However, for glasses that are quenched rapidly to ambient temperature and hydrated subsequently, it is expected that the OH content of a hydrated sample should remain the same as when that sample was first deposited. This is in keeping with observations of volcaniclastic glasses that contain hydrated regions. Those hydrated regions have elevated H2Om concentrations, but have similar OH concentrations to unhydrated regions of the same samples (e.g., Nichols et al. 2014). By using the new species-dependent ε3500 method to measure accurately the OH concentration of such hydrated glasses, it is now possible to estimate the original pre-hydration H2Ot content by using speciation models (e.g., Nowak and Behrens 2001; Fig. 6) to find the H2Ot concentration that corresponds to the measured OH concentration for the expected Tae (≈ Tg) of the sample (e.g., Dingwell et al. 2015). Although the glass transition temperature can vary with cooling history and H2Ot concentration, OH vs. H2Ot curves for different values of Tg converge at low-H2Ot concentrations (Fig. 6), making this an effective method for glasses with <1 wt% OH. Other proposed methodologies for reconstructing the original H2Ot content of hydrated glasses involve thermogravimetric analysis (TGA) (e.g., Denton et al. 2009, 2012; Tuffen et al. 2010; Giachetti et al. 2015) or hydrogen isotope analysis (e.g., DeGroat-Nelson et al. 2001)—both of which produce bulk measurements and destroy the sample—in conjunction with modeling of diffusion, for which some of the parameters are not well constrained for ambient temperatures. This FTIR methodology is relatively cheap and simple to perform, and has the significant benefit of permitting spatial variations in both the original and subsequent H2Ot concentration to be measured.

Compositional dependence of the molar absorptivity coefficients

It has long been recognized that the values of FTIR molar absorptivity coefficients vary with glass composition (e.g., Silver et al. 1990; Dixon et al. 1995; Ohlhorst et al. 2001; Mandeville et al. 2002; Seaman et al. 2009; Mercier et al. 2010), hence it is unsurprising that the end-member ε3500 values also vary with glass composition. H2Ot and OH contents derived using the rhyolite end-member coefficients agree well with the manometry data for both of the rhyolite glass standards (Fig. 4), suggesting that these end-member ε3500 values are not sensitive to minor compositional differences, such as a few wt% of SiO2 (Table 3), and can be successfully applied to other glasses with similar major element compositions. However, glasses with greater compositional differences will require their own set of end-member coefficients, as seen in the variation between values for rhyolite, albite, and Fe-bearing and Fe-free andesite (Table 2).

Previous studies have suggested that the molar absorptivity coefficients of the H2O absorption bands decrease with decreasing tetrahedral cation fraction (τ) of the melt, where τ = (Si4++Al3+)/total cations (e.g., Dixon et al. 1995; Ohlhorst et al. 2001; Mandeville et al. 2002; Seaman et al. 2009; Mercier et al. 2010). Although our data are so far limited to only four glass compositions, our values of ε3500OH and ε3500H2Om do not show a trend with τ (Table 2), and neither do they show a trend with the ratio of non-bridging O atoms over tetrahedrally coordinated cations (NBO/T; Table 2). Although the end-member ε3500 values clearly vary with melt composition, it is not yet possible to simply link this variation with a particular structural parameter that describes the silicate melt.

Of the four compositions discussed here, the use of a species-dependent ε3500 is most important for the rhyolite and Fe-andesite compositions. These compositions have the greatest difference between the two end-member coefficients (56–100 and 49–79, respectively) hence the appropriate ε3500 value, and the calculated H2Ot and OH concentrations, can vary widely. For Fe-free andesite the difference between the end-member coefficients is smaller (62–76) but is still sufficient to justify the use of species-dependent ε3500. On the other hand, the difference between end-member ε3500 values for albite is so small (69–71) that they are within error of each other, which enabled Silver and Stolper (1989) to conclude that there is no advantage in choosing a species-dependent ε3500 value over a fixed ε3500 value for this composition.

Dissolved H2O content exerts a strong control on silicate melt properties such as viscosity, glass transition temperature, diffusivities of mobile species, and crystallization kinetics. Consequently, for many studies throughout the geological, archeological, and materials sciences, it is critical to know accurately the H2O contents of silicate glasses, even when the behavior of H2O is not the main focus of the study. The relative ease of sample preparation and low cost of FTIR analyses means that, in many studies of this sort, the H2O data are obtained by FTIR, and often by analyzing the 3500 cm–1 H2Ot absorbance band. Our new species-dependent ε3500 method, which does not require a change in analytical procedure or expensive instrumentation, will improve the accuracy of these FTIR data, and remove systematic bias in samples that have undergone hydration. Previously published H2O data that depend upon fixed ε3500 values will therefore need to be re-evaluated to identify and correct potential inaccuracies. This is particularly important given the common use of FTIR H2O data to constrain glass standards used in the calibration of other analytical techniques such as SIMS (e.g., Hauri et al. 2002). We emphasize that this correction can be accomplished without the need for new instrumental analyses, by simply re-analyzing the raw absorbance, thickness, and density data that have already been collected, using the methodology developed in this work. We provide a simple spreadsheet, which accepts these data and performs the relevant calculations, in the supplementary information1.

The clear advantage of FTIR analysis over other techniques is the ability to obtain quantitative H2O speciation data, which can be used to investigate topics such as the mechanisms and rates of H2O diffusion (e.g., Zhang et al. 1991), or the pressure and temperature histories of glasses (e.g., Zhang et al. 1997a; McIntosh et al. 2014). Many of these studies involve measuring the spatial variation in H2O species along an H2O diffusion profile, for example along a diffusion couple experiment or toward a crack, bubble or crystal in the glass (e.g., Zhang et al. 1991; Castro et al. 2008; Berlo et al. 2013; von Aulock et al. 2013; McIntosh et al. 2014; Saubin et al. 2016; Watkins et al. 2017). In such cases, where H2Ot concentration varies systematically with spatial position, it is essential to account for the species-dependence of ε3500 when using the 3500 cm–1 absorbance band, because otherwise errors will also vary systematically with spatial position, as illustrated by our reinterpretation of the Yokoyama et al. hydration profiles (Fig. 7). Removal of this systematic error will significantly improve the quality of interpretations of H2O speciation data along such profiles.

As well as facilitating reanalysis and reinterpretation of existing studies of experimental and natural hydrous glasses, our new method also opens new avenues of research into glasses affected by secondary hydration. Secondary hydration has been shown to be a widespread phenomenon (e.g., Giachetti and Gonnermann 2013), and we recommend the use of FTIR analyses to identify (by their excess H2Om contents) glasses that have been hydrated; information that cannot be obtained by SIMS or Raman spectroscopy. Moreover, improved accuracy of H2O speciation data along hydration profiles will benefit researchers investigating mechanisms of glass hydration, and the use of obsidian hydration profiles as a dating tool for volcanic and archeological glasses. Most exciting of all, however, is the prospect of estimating the pre-hydration H2Ot contents of hydrated glasses based on accurate measurement of their OH contents. We propose that this will be a particularly important breakthrough for understanding the eruption processes and associated hazards of silicic submarine eruptions, for which pyroclast glasses are routinely found to be hydrated (e.g., Bryant et al. 2003; Kutterolf et al. 2014). In the same way that the pressure-dependence of H2O solubility can be used to determine paleo-ice thicknesses from subglacially erupted and quenched glasses (e.g., Tuffen et al. 2010), reconstructed H2Ot contents of submarine pyroclast glasses can be used to establish the quench depth of unobserved submarine eruption plumes and, in particular, to determine whether pyroclasts now resting on the deep sea floor may have originally reached the sea surface (e.g., Fiske et al. 2001; Tani et al. 2008; Allen et al. 2010; Rotella et al. 2013).

Although this paper has focused necessarily on rhyolitic and andesitic glasses, it is hoped that future FTIR studies of hydrous silicate glasses will enable further sets of end-member ε3500 values to be derived for more compositions (particularly basalt). This will allow the species-dependent ε3500 method to be applied more widely, and will also improve our understanding of the composition dependence of molar absorptivity coefficients. Such future studies may also improve upon the molar absorptivity coefficients that are available to us today. This study has benefited hugely from the previously published data sets of Newman et al. (1986) and Mandeville et al. (2002), which contained the absorbance, thickness, and density data necessary to test and develop our methodology. We therefore strongly advocate for researchers using FTIR data to publish, as we have done here, not only their concentration data but also the absorbance, thickness and density data that underpin them, with the aim of increasing the longevity and relevance of their hard-won data.

1Deposit item AM-17-85952, Supplemental Material: A Microsoft Excel spreadsheet for using the species-dependent ε3500 method and for deriving new end-member ε3500 values from absorbance data. Deposit items are free to all readers and found on the MSA web site, via the specific issue's Table of Contents (go to http://www.minsocam.org/MSA/AmMin/TOC/2017/Aug2017_data/Aug2017_data.html).

We are grateful to Madeleine Humphreys for providing the glass standards used in this study. We thank Jake Lowenstern and an anonymous reviewer for their thorough and insightful reviews. This research was supported by a Japan Society for the Promotion of Science Postdoctoral Fellowship for Foreign Researchers held by I.M., and a JSPS Grant-in-Aid for Scientific Research (Kakenhi grant number 00470120) awarded to A.R.L.N and I.M.; E.W.L. acknowledges support from the U.K. Natural Environment Research Council via grant NE/N002954/1. Data Access Statement: Underlying data can be obtained by contacting the corresponding author.